Categories
1. Preanalytical variables & experimental design
1.1 What should be reported:
Preanalytical variables relating to EV sample including source, collection, isolation, storage, and any others relevant and available in the performed study.
Reported:
-
1.2 What should be reported:
EV-FC manuscripts should provide a brief description of the experimental aim, keywords, and variables for the performed FC experiment(s) using MIFlowCyt checklist criteria: 1.1, , and 1.3, respectively. Template found at www.evflowcytometry.org.
Reported:
See 'MIFlowCyt' sheet.
2. Sample preparation
2.1 What should be reported:
State any steps relating to the staining of samples. Along with the method used for staining, provide relevant reagent descriptions as listed in MIFlowCyt guidelines (Section 2.4 Fluorescence Reagent(s) Descriptions).
Reported:
-
2.2 What should be reported:
State any steps relating to the washing of samples.
Reported:
-
2.3 What should be reported:
All methods and steps relating to sample dilution.
Reported:
-
3. Assay controls
3.1 What should be reported:
State whether a buffer-only control was analyzed at the same settings and during the same experiment as the samples of interest. If utilized it is recommended that all samples be recorded for a consistent set period of time e.g. 5 minutes, rather than stopping analysis at a set recorded event count e.g. 100,000 events. This allows comparisons of total particle counts between controls and samples.
Reported:
-
3.2 What should be reported:
State whether a buffer with reagent control was analyzed at the same settings, same concentrations, and during the same experiment as the samples of interest. If used state what the results were.
Reported:
-
3.3 What should be reported:
State whether unstained control samples were analyzed at the same settings and during the same experiment as stained samples. If used, state what the results were, preferably in standard units.
Reported:
-
3.4 What should be reported:
The use of isotype controls is applicable to immunofluorescence labelling only. State whether isotype controls were analyzed at the same settings and during the same experiment as stained samples. If utilized, state which antibody they are matched to, the concentration used, and what the results were (Section 4.2, 4.3, 4.4). Due to conjugation differences between manufacturers if should be stated if the isotype controls are from the same manufacturer as the matched antibodies.
Reported:
-
3.5 What should be reported:
State whether single-stained controls were included. If used state whether the single-stained controls were recorded using the same settings, dilutions, and during the same experiment as stained samples and state what the results were, preferably in standard units (Section 4.2, 4.3, 4.4).
Reported:
-
3.6 What should be reported:
State whether procedural controls were included. If used, state the procedure and if the procedural controls were acquired at the same settings and during the same experiment as stained samples.
Reported:
-
3.7 What should be reported:
State whether serial dilutions were performed on samples and note the dilution range and manner of testing. The fluorescence and/or scatter signal intensity would ideally be reported in standard units (see Section 4.3, 4.4) but arbitrary units can also be used. This data is best reported by plotting the recorded number events/concentration over a set period of time at different sample dilution. The median fluorescence intensity at each of the dilutions should also ideally be plotted on the same or a separate plot.
Reported:
-
3.8. What should be reported:
State whether samples were detergent treated to assess lability. If utilized, state what detergent was used, the end concentration of the detergent, and what the results were of the lysis.
Reported:
-
4. Instrument calibration & data acquisition
4.1 What should be reported:
The trigger channel(s) and threshold(s) used for event detection. Preferably, the fluorescence calibration (Section 4.3) and/or scatter calibration (Section 4.4) should be used in order to report the trigger channel(s) and threshold(s) in standardized units.
Reported:
Trigger Channel(s) and Threshold(s) in calibrated and arbitrary units are summarized for each sample in the table on the 'Sample Acq Info' sheet.
4.2 What should be reported:
State if the flow rate was quantified/validated and if so, report the result and how they were obtained.
Reported:
-
4.3 What should be reported:
State whether fluorescence calibration was implemented, and if so, report the materials and methods used, catalogue numbers, lot numbers, and supplied reference units for the standards. Fluorescence parameters may be reported in standardized units of MESF, ERF, or ABC beads. The type of regression used, and the resulting scatter plot of arbitrary data vs standard data for the reference particles should be supplied.
Reported:
Fluorescence calibration was performed on 1 parameter(s) using FCMPASS software as previously demonstrated [1,2,3]. Details on the calibration reagents and regression can be found in the summary table on the 'Fl Cali' sheet and in FCMPASS Figure 2.
4.4 What should be reported:
State whether and how light scatter calibration was implemented. Light scatter parameters may be reported in standardized units of nm2, along with information required to reproduce the model.
Reported:
Light scatter calibration was performed using FCMPASS software as previously demonstrated [1,2,3]. Details on the reagents and modelling parameters can be found in the summary table on the 'SSC Cali' sheet and in FCMPASS Figure 1.
5. EV characterization
5.1 What should be reported:
State whether and how EV diameter, surface area, and/or volume has been calculated using FC measurements.
Reported:
Diameter approximation using light scattering was performed using FCMPASS software as previously demonstrated [1,2,3]. Details of modelling parameters for calibration, core-shell, and solid sphere models can be found in the 'SSC Cali' sheet and in FCMPASS Figure 1.
5.2 What should be reported:
State whether the EV refractive index has been approximated and how this was done.
Reported:
-
5.3 What should be reported:
State whether EV epitope number has been approximated, and if so, how it was approximated.
Reported:
-
6. FC data reporting
6.1 What should be reported:
Complete MIFlowCyt checklist criteria 1 to 4 using the MIFlowCyt guidelines. Template found at www.evflowcytometry.org.
Reported:
See 'MIFlowCyt' sheet.
6.2 What should be reported:
If fluorescence or scatter calibration has been carried out, authors should state whether the upper and lower limits of a calibrated detection channel were calculated in standardized units. This can be done by converting the arbitrary unit scale to a calibrated scaled, as discussed in Section 4.3 and 4.4, and providing the highest unit on this scale and the lowest detectable unit above the unstained population. The lowest unit at which a population is deemed ‘positive’ can be determined a variety of ways, including reporting the 99th percentile measurement unit of the unstained population for fluorescence. The chosen method for determining at what unit an event was deemed positive should be clearly outlined.
Reported:
-
6.3 What should be reported:
State whether EV number/concentration has been reported. If calculated, it is preferable to report EV number/concentration in a standardized manner, stating the number/concentration between a set detection range.
Reported:
-
6.4 What should be reported:
When applicable, state the method by which the brightness of EVs is reported in standardized units of scatter and/or fluorescence.
Reported:
-
7. FC data sharing
7.1. What should be reported:
Provide a link to the experimental data in a public data repository.